Call Us: 844-742-3818 | Email Us: info@DiscSci.com
Model | CR22N |
Max. Speed (rpm) | 22,000 rpm |
Max. RCF | 55,200 xg (R22A4) |
Max. Capacity | 1,500 ml x 4 bottles (R9A2) |
Speed Control Range | 300 ~ 22,000 rpm |
Temperature Control Range | -20 ~+40° C |
ACL/DCL Mode | ACL: 9 modes, DCL: 9 modes with coasting |
Program Function | 30 programs, with RCF(g) / g x sec function |
Drive Unit | Induction Motor (Inverter) |
Refrigerator | Closed Inverter Rotary (Refrigerant : R410A) |
Safety Functions | Interlock Door, Dual Over Speed Detection (Non-contact Imbalance Detection, Abnormal Temp. Detection etc.) |
Variable Deceleration Slope | Yes (beside DCL mode , moderate deceleration or free coast is selectable.) |
User Lockout Function | Yes |
Data Communication | USB port (Operating history can be output through USB port.) |
Operating Log Management | LAN : Factory Option PC and optional himac LogManager Ver.3.1 network edition is required. Supporting GLP/GMP |
Display | High-resolution 5.7 inch color LCD |
Keypad | Touch-sensitive LCD panel (with GUI icons) |
Noise | 64 dB(A) at 1m front (at 22,000rpm) |
Economy Mode | Yes |
Standard | ??? |
(1) EMC Directive | EN61326-1 |
(2) LVD and MD* | EN61010-1, EN61010-2-020 |
Dimensions (W x D x H : mm) | 700 x 810 x 915 (height to chamber : 853) |
Weight | 290 kg |
Power Requirements | Single Phase , AC 200, 208, 220, 230, 240V+/-10%, 30A (50/60Hz) |
Note : 1. Please note that LCD panel may contain a few dead or stuck pixels. 2. *LVD : Low Voltage Directive, MD : Machinery Directive