Call Us: 844-742-3818 | Email Us: info@DiscSci.com
Hypoxic Zirconia Ceramic O2 Sensor Control
A maintenance-free ZirconiaO2 ceramic sensor provides improved accuracy and lower range hypoxic control by displacement with nitrogen.
Example Studies:
O2 Range: 0.5 to 21%
O2 Accuracy: ± 0.25%
Dual Sterilization Cycles
Easily accessible through the NuTouch Intelligent Interface, In-VitroCell offers 145°C high heat sterilization and a 95°C humidified decontamination cycles as a preventative measures and to eradicate contamination in the growth chamber. On screen instructions through NuTouch guide users through the process. A timer function is available to delay the sterilization process for an overnight procedure.
NuTouch Intelligent Interface
NuTouch is a user-friendly 5 x 7 (127 x 178 mm) color touch screen designed for high user adoption. The electronic control system is a microcomputer designed to service the precise control requirement of the chamber environment for culture growth. The microcomputer is supported with Read Only Memory (ROM) containing executable software, Random Access Memory (RAM) for temporary storage, and Electronically Erasable Programmable Read Only Memory (EEPROM) for control set points and parameters. The EEPROM provides for indefinite storage of these values during periods of power off or power interruption (power fault tolerant). NuTouch features are:
Constant Contamination Control (C3)
A unique blend of product features to promote cell growth while minimizing potential contamination creating the most intrinsic and reliable growth chamber.
Closed Loop HEPA Filtration
A maintenance free air pump continuously draws environmental samples from two locations inside the chamber and passes the sample through a 99.99% HEPA filter before entering the sensor bay to measure gas and oxygen levels. The IR sensor measures CO2 gas while a Zirconia Cermanic O2 sensor measures oxygen levels from the sample to monitor before injecting back into the growth chamber. The NuTouch Electronic Control system will then determine the appropriate course of action to maintain chamber set-point conditions by placing the system on standby or injecting 99.99% HEPA filtered CO2 gas, air, or nitrogen into the growth environment.
ISO Class 5 Cleanroom Conditions
Based on clean room technology, In-VitroCell maintains the growth chamber at positive pressure similar to an ISO Class 5 Clean room to slow airflow to minimize desiccation. Anytime the chamber door is open to the laboratory environment, clean HEPA filtered will be forced out instead of contaminated lab air being drawn in.
Single Beam, Dual Wave Infrared (IR) Sensor
A microprocessor-based, non-dispersive, single source dual wave infrared (IR) sensor monitors and controls CO2 levels inside the growth chamber. Wavelengths are only absorbed by CO2 making measurement insensitive to other components, such as water vapor. The stable IR sensor has a controlled range of 0.1 to 20% accurate within 0.1%.
Temperature Uniformity
The chamber walls are surrounded by five (5) heating elements to control temperature. A door heater is separately controlled to minimize condensation on the inner chamber door. The heating elements are wrapped in a high density R5 insulation to maintain temperature integrity. Dual temperature sensor probes monitor and control temperature set points to ensure consistency and uniformity throughout the chamber.
Incubator Humidity
A relative humidity level of up to 90% is achieved in the Incubator by the use of a stainless steel pan placed on the bottom of the chamber. An air pump injects fresh air into the chamber at a preset duty cycle to reduce condensation forming on the chamber walls and front glass door.
Advanced Construction
The inner chamber is constructed of high grade 16 gauge, type 304L polished stainless steel using crevice-free construction that provides an easily cleanable inert surface that does NOT promote biological growth. Seamless coved interior corners are easily cleaned promoting a healthy growth chamber. A single piece gasket creates a liquid tight seal from incubator chamber to inner door to eliminate condensation build up and potential contamination. Inner chamber shelving and gasket may be removed and routinely autoclaved if desired.
Minimal Vibration Promotes Cell Growth
By minimizes moving parts, In-VitroCell ES utilizes an accurate air pump to inject air and gas to eliminate potential vibration.
Smaller Footprint with More Useable Space
A large 7 cubic foot (200 Litre) growth chamber comes standard with four incubator shelves with a maximum capacity of 23 shelves. By attaching heating elements directly to the chamber and insulating with High Density R5 insulation In-VitroCell was engineered as a compact design while maximizing chamber capacity.
CuVerro® Antimicrobial Copper Surfaces (Optional Feature)
Add CuVerro® Antimicrobial Copper Surfaces to the incubator growth chamber and/or shelving to kill bacteria* to minimize potential incubator contamination. CuVerro® is laboratory tested and EPA registered. CuVerro® Antimicrobial Copper Surfaces kill more than 99.9% of bacteria* within 2 hours, and continues to kill 99% of bacteria* even after repeated contamination, when cleaned regularly.
EPA Reg No 85353-5
EPA Est No 088257-MN-001
*Laboratory testing shows that, when cleaned regularly, CuVerro® antimicrobial copper surfaces kill greater than 99.9% of the following bacteria within 2 hours of exposure: MRSA, Staphylococcus aureus, Enterobacter aerogenes, Pseudomonas aeruginosa, and E. coli O157:H7. CuVerro® antimicrobial copper surfaces are a supplement to and not a substitute for standard infection control practices and have been shown to reduce microbial contamination, but do not necessarily prevent cross contamination; users must continue to follow all current infection control practices, including those practices related to cleaning and disinfection of environmental surfaces.
Standard features
Four (4) Stainless Steel Shelves
Eight (8) Stainless Steel
Shelf Brackets
Shelf Guides
Right Hinge Door
Remote Alarm Output Contacts
4 to 20 mA Analog Output
RS-485 Communication
USB Port
CO2 Sample Port
Adjustable Leg Levelers
Access Port and plug with breather holes
One (1) Water Pan
One (1) 6.5 ft. / 2 m Electrical Cord
Optional Features
Automatic CO2 Tank Switch (External)
CO2 Tank Alarm
Left Hinge Door
Additional Shelves with Slide Brackets
CO2 Analyzer Fyrite Kit (Dry) 0-20%
Replacement Fluid for CO2
Analyzer
Surge Protector
CO2 Regulator (2 Stage)
Custom Solutions