Description
Space-saving design with lid opening upward.
Fully automatic drying cycle for solid loads uses heat and vacuum for faster and more efficient drying.
Lid safety interlock prevents lid opening when a cycle is in progress, the chamber is pressurized, or when the temperature exceeds 97°C.
Pre-vacuum cycle - vacuum pump removes air from the chamber when sterilizing solids such as glassware and pipette tips, thereby eliminating cold spots, and ensuring faster and more efficient sterilization.
One-touch lever to easily open and close chamber lid.
Liquid cycle for media and agar.
Process display shows the program progress selected and status of the sterilization cycle.
Cycle for wastes and glassware without drying.
Bacteria filter reduces re-contamination.
Door-closure sensor prevents a cycle from starting unless the door is closed properly.
HVP-50 size: 50 liters, 11.8" D, 28.0" H.
WARRANTY: Autoclaves are protected by a 12-month warranty.
Specifications
Amerex HVP-50 Autoclave Specifications
TEMPERATURE |
Operating Range |
105ºC to 135ºC |
Display Range |
5ºC to 140ºC |
GAUGE PRESSURE |
Operating Range (gauge) |
3 to 28.4 psi (0.2 to 2.0 kg/cm² ) |
TIMER |
Sterilization |
1 to 250 minutes |
Drying |
1 to 90 minutes |
DIMENSIONS |
Overall (W x D x H) |
20.1" x 24.8" x 40.9" (51 x 63 x 104) cm |
Chamber / (Diameter x Height) |
50 liter / 11.8" x 28" (30 cm x 71 cm) |
SAFETY DEVICES |
Temperature- and pressure-sensitive lid interlock, over-pressure power cutoff, over-temperature power cutoff, low-water power cutoff, mechanical pressure relief valve, lid-closure detector, over-current detector and circuit breaker with ground fault circuit interrupter. |
PROGRAMS |
1. For solids with drying |
Heating and air removal from chamber by vacuum pump (pre-vac) -> Sterilization -> Water and steam exhaust (fast) -> Drying by heat and vacuum -> Completion |
2. For wastes/solids without drying |
Heating -> Pre-vac -> Sterilization -> Steam exhaust (fast) -> Completion |
3. For liquids/agar |
Heating -> Sterilization -> Auto steam exhaust (programmable rates, or ambient cooling w/o steam exhaust) -> Completion |
4. Drying without sterilization |
Drying with heat and vacuum (time programmable) -> Completion(for additional drying after Program #1, if desired) |
ACCESSORIES |
Stainless steel |
Mesh wire baskets, buckets (pails), and round test tube racks |
Factory or field installed |
Data logger |
POWER |
Source |
120V, 50/60 Hz, single phase (220V/240V available on request) |
Consumption |
2000 watts (2000 VA) |
SHIPPING WEIGHT |
291 lb (133 kg) |